Télécharger Imprimer la page

Hettich ROTO SILENTA 630 RS Mode D'emploi page 71

Masquer les pouces Voir aussi pour ROTO SILENTA 630 RS:

Publicité

Les langues disponibles

Les langues disponibles

Fig. 10: [STOP] indicator
3.5.3
Controls
Fig. 11: [Rotary knob]
Fig. 12: [Mains switch]
Fig. 13: [Key switch]
Fig. 14: [Temperature and
centrifuging radius] button
Fig. 15: [Ramp-up parameters]
button
Fig. 16: [Ramp-down
parameters] button
AB5005en
The indicator appears during the centrifugation run as long as the rotor
is turning.
The indicator flashes after an emergency stop.
The indicator flashes after an emergency stop.
Setting the individual parameters.
Turning anticlockwise decreases the value.
Turning clockwise increases the value.
Switch the device on and off.
The key switch switches various functions on and off, depending on the
position.
Temperature setpoint, parameter T/°C
Adjustable from -20°C to +40°C, in 1°C increments (adjustable from
-20°C to +90°C with heating/cooling option).
The lowest achievable temperature is rotor dependent.
Centrifuging radius
Parameter r/mm. Input in mm.
Ramp-up levels, parameters
Level 9 = shortest ramp-up time, Level 1 = longest ramp-up time.
Ramp-up time parameters
The adjustable time range is dependent on the set speed.
Brake levels, parameters
R = Linear braking curve,
B = similar to an exponential braking curve.
Level R9, B9 = short ramp-down time, ...
Level R1, B1 = long ramp-down time,
Level R0 = unbraked ramp-down.
Ramp-down time, parameters
The adjustable time range is dependent on the set speed.
Brake cut-off speed, parameter n
Unbraked ramp-down takes place after reaching this speed.
Rev.: 20 / 01.2024
Device overview
 /RPM
(·)
17 / 50

Publicité

loading