Software operation
7.3.3
Automatic buffer
7.4
Rotor detection
7.5
Cooling (for centrifuges with cooling)
7.5.1
Instructions, cooling
7.5.2
Standby cooling
7.5.3
Precooling the rotor
7.6
Heating (for centrifuges with heating)
36 / 50
[PROG] button to select the 'PROG no.' parameter. The input
2.
Use the
field is reverse highlighted.
[Rotary knob] to set the desired program location.
3.
Use the
If the program location indicator flashes, this program location is
already assigned centrifugation data. In this case, set a free program
location, or overwrite the centrifugation data by continuing.
4.
Press the
[STO] button.
Settings are stored in the desired program location.
[STO] button twice.
5.
Press the
Centrifugation data that is already stored will be overwritten.
The buffer includes the program locations "----" and 90 to 99.
Changed centrifugation data is automatically saved to program location
"----" every time a centrifugation run is started.
The changed centrifugation data, of the last 11 centrifugation runs, are
stored in the buffer and can be retrieved.
■
Rotor detection is performed after starting a centrifugation run.
■
If the rotor has been changed, the centrifugation run is cancelled after
rotor detection. The rotor code (R) and the maximum rotor speed (n-
max) of the newly detected rotor are displayed.
■
If the maximum speed of the rotor used is less than the set speed, the
speed is limited to the maximum rotor speed.
For centrifuges with the heating/cooling option, the temperature setpoint
can be adjusted from -20 °C to +90 °C. If the actual temperature deviates
from the set temperature by more than 5 °C, this is signalled by a flashing
temperature value indicator.
The lowest achievable temperature is rotor dependent.
The centrifuging chamber is cooled to the preselected temperature when
the rotor is at a standstill and the lid is closed. The display shows the
temperature setpoint.
For rapid precooling of the unloaded rotor and accessories, we recommend
a centrifugation run with the continuous operation settings and a speed of
approx. 20% of the maximum rotor speed.
During the centrifugation run, the centrifuging chamber is heated to the
preselected temperature if required. The heating is switched off when the
rotor is at a standstill.
Rev.: 20 / 01.2024
AB5005en