Software operation
Changing a program link
7.8.2
Centrifugation run with program link
40 / 50
[STO] button twice.
5.
Press the
The program is linked and the program number of the next pro-
gram location (+XX+) is displayed.
[RCL] button twice.
6.
Press the
The centrifugation data of the desired program location is dis-
played
7.
Press the
[STO] button twice.
The program is linked and the program number of the next pro-
gram location (+XX+) is displayed.
8.
Repeat the last two steps until all programs are linked.
[PROG] button.
9.
Press the
The program number of the end program (+XX) is displayed.
1.
Open the desired program.
2.
Change the desired parameter.
3.
Save the changed centrifugation data to the same program location
again.
Saving removes the program link.
4.
Link programs again.
[PROG] button twice.
1.
Press the
Parameter PR-PART is selected.
The input field is reverse highlighted.
[Rotary knob] to set the program location of the initial pro-
2.
Use the
gram (XX+).
[RCL] button.
3.
Press the
The centrifugation data of the desired program location is dis-
played
4.
Press the
[START] button.
The centrifugation run is started.
'Rotation' indicator appears as long as the rotor is turning.
The
The ramp-up and brake levels of the program link are displayed.
■
Initial program (XX+)
x: Ramp-up level of the initial program x
■
Follow-up program (+XX+)
x: Ramp-up level of the follow-up program x
■
End program (+XX)
x: Ramp-up level of the end program
y: Brake level of the end program
5.
Ramp-down takes place with the brake level of the end program after
the time in the end program has elapsed.
If the centrifugation run is cancelled by pressing the
ramp-down takes place at the brake level of the program currently
running.
Rev.: 20 / 01.2024
[STOP] button,
AB5005en