Troubleshooting
Fault description
N < MIN 13
ROTORCODE 10
MAINS INTERRUPT
VERSION-ERROR 12
SER I/O - ERROR 30-38
° C * - ERROR 50-56, 58
LOCK - ERROR 57
FU/CCI-ERROR 60-83
CONTROL - ERROR 26,
90-95, 97 - 99
N > ROTOR MAX 96
The entire
display lights up.
9.2
Performing a MAINS RESET
9.3
Emergency release
46 / 50
Cause
Underspeed error.
Rotor coding error.
Loss of mains power during the
centrifugation run. The centrifuga-
tion run was not completed.
No conformity of the electronic
components, error/defect in elec-
tronics.
Error/defect in interface.
Error/defect in cooling.
Error/defect in program lock.
Error/defect in motor control.
Error/defect in control section.
Speed in the selected program
greater than the maximum rotor
speed.
The rotor has been changed. The
built-in rotor has a higher maximum
speed than the previously used
rotor. The rotor has not yet been
recognised by the rotor detection.
-
1.
Set the mains switch to
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
−
Disconnect the device from the mains before carrying
out repairs and maintenance.
Rev.: 20 / 01.2024
Remedy
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Open the lid.
■
Press the
[START] button.
■
If required: Repeat the centrifuga-
tion run.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Perform a MAINS RESET.
■
Check and correct the speed.
■
Set a speed up to the maximum
speed of the previously used
rotor. Press the
perform rotor detection.
■
Notify customer service.
[0] .
[I] .
[START] button to