Télécharger Imprimer la page

Hettich ROTO SILENTA 630 RS Mode D'emploi page 100

Masquer les pouces Voir aussi pour ROTO SILENTA 630 RS:

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting
Fault description
N < MIN 13
ROTORCODE 10
MAINS INTERRUPT
VERSION-ERROR 12
SER I/O - ERROR 30-38
° C * - ERROR 50-56, 58
LOCK - ERROR 57
FU/CCI-ERROR 60-83
CONTROL - ERROR 26,
90-95, 97 - 99
N > ROTOR MAX 96
The entire
display lights up.
9.2
Performing a MAINS RESET
9.3
Emergency release
46 / 50
Cause
Underspeed error.
Rotor coding error.
Loss of mains power during the
centrifugation run. The centrifuga-
tion run was not completed.
No conformity of the electronic
components, error/defect in elec-
tronics.
Error/defect in interface.
Error/defect in cooling.
Error/defect in program lock.
Error/defect in motor control.
Error/defect in control section.
Speed in the selected program
greater than the maximum rotor
speed.
The rotor has been changed. The
built-in rotor has a higher maximum
speed than the previously used
rotor. The rotor has not yet been
recognised by the rotor detection.
-
1.
Set the mains switch to
2.
Wait 10 seconds.
3.
Set the mains switch to
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
Disconnect the device from the mains before carrying
out repairs and maintenance.
Rev.: 20 / 01.2024
Remedy
Perform a MAINS RESET.
Perform a MAINS RESET.
Open the lid.
Press the
[START] button.
If required: Repeat the centrifuga-
tion run.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Perform a MAINS RESET.
Check and correct the speed.
Set a speed up to the maximum
speed of the previously used
rotor. Press the
perform rotor detection.
Notify customer service.
[0] .
[I] .
[START] button to
AB5005en

Publicité

loading