Commissioning
5.2
Setting up and connecting the centrifuge
Setting up the centrifuge
Connecting the centrifuge
22 / 50
7.
Push the centrifuge to its installation location.
8.
Place the open-end spanner (size 10 mm) on the surfaces (
2 ) down until the castors are no longer in contact with
turn the feet (
the ground.
9.
Align the centrifuge so it is horizontal by turning the device feet (
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
−
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
−
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
−
Place the device on a stable and level surface.
−
Select the installation surface dependent on the weight
of the device.
NOTICE
Damage to the samples and the device if the ambient
temperature exceeds or falls below the respective max-
imum/minimum permissible ambient temperature.
−
Comply with the maximum and minimum permissible
ambient temperatures for installation of the device.
−
Do not place the device next to a heat source.
−
Do not expose the device to direct sunlight.
−
Do not expose the device to frost.
Personnel:
■
Trained user
1.
Place the device on a stable and level surface.
2.
Maintain a distance of 300 mm around the device.
3.
Comply with the ambient conditions in the technical data (⮫
3.1 'Technical data' on page 10 ).
NOTICE
Damage to the device by unauthorised personnel
−
Tampering with and modifications to devices by unau-
thorised persons are at the operating organisation's own
risk and will result in the loss of all warranty and liability
claims.
Rev.: 20 / 01.2024
1 ) and
2 ).
Chapter
AB5005en