Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
Ambient conditions (EN / IEC 61010-1):
Installation site
Altitude
Ambient temperature
Humidity
Overvoltage category
(IEC 60364-4-443)
Pollution level
Device protection class
EMC:
Emitted EM interference,
EM interference immunity
Noise level
(rotor-dependent)
Dimensions:
Width
Depth
Altitude
Weight
Manufacturer
Model
Type
Mains voltage (±10%)
Mains frequency
power consumption
Power consumption
AB5005en
yes
indoors only
up to 2000 m above sea level
5 °C to 40 °C
maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C.
II
2
I
not suitable for use in potentially explosive atmospheres.
EN / IEC 61326-1
Class B
£62 dB(A)
813 mm
1015 mm
973 mm
approx. 355 kg
Andreas Hettich GmbH & Co. KG,
D-78532 Tuttlingen
ROTO SILENTA 630 RS
5005-08
208-220  V +6/-10%  3~ (+N) +PE
50-60 Hz
9000 VA
25 A
Rev.: 20 / 01.2024
1050 mm
approx. 367 kg
Device overview
£56 dB(A)
approx. 306 kg
11 / 50