Software operation
7.7.2
Querying operating hours
7.7.3
Audible signal
7.7.3.1
General
7.7.3.2
Setting an audible signal
7.7.4
Centrifugation data displayed after switching on
38 / 50
The rotor is stationary.
1.
Open the lid.
2.
Press and hold the
'SOUND / BELL XXX' is displayed after 8 seconds.
[t] button.
3.
Press the
'CONTROL:' and the operating hours are displayed.
The operating hours indicator goes off automatically after 10 sec-
onds.
The audible signal sounds after the following settings:
■
OFF
after a problem occurs in the 2 s interval.
■
ON1
after a problem occurs in the 2 s interval.
■
after completion of the centrifugation run and rotor
standstill in the 30 s interval.
■
ON2
after a problem occurs in the 2 s interval.
■
after completion of the centrifugation run and rotor
standstill in the 30 s interval.
■
every time the button is pressed.
Opening the lid or pressing any button stops the audible signal.
1.
Open the lid.
2.
Press and hold the
'SOUND / BELL ON1' , 'SOUND / BELL ON2' or 'SOUND / BELL
OFF' is displayed after 8 seconds.
[Rotary knob] to set 'OFF' , 'ON1' or 'ON2' .
3.
Use
[START] button.
4.
Press the
The setting is stored.
'*** OK ***' is displayed briefly.
The centrifugation data of program 1 or the last program used is displayed
after switching on.
1.
Set the mains switch to
[STOP] button at the first visual change in the display
2.
Press the
(inverse display).
'PROGRAM 1, LAST PROGRAM' is displayed.
[Rotary knob] to set the desired function.
3.
Use the
Rev.: 20 / 01.2024
[t] button.
[t] button.
[I] .