Télécharger Imprimer la page

Leica MATEO TL Instructions D'utilisation page 9

Publicité

Les langues disponibles

Les langues disponibles

Electrical safety
Use only original power cables or alternative
cables with a VDE / HAR logo, which at least fulfill the requirement of
3x0.75 mm
and 10A/250V. Use the original power supply only (LPS-
2
certified Power Supply with the same specifications).
Make sure that the power cord is approved for use in the country
in which you intend to use it.
The cables shall be plugged or unplugged in the de-energized
state! Before connecting the system, check that the supply voltage and
frequency are correct at the installation site.
Always hold the plug of power supply when removing it from
the socket outlet. Never unplug it by pulling the cable.
If the original power supply fails or is damaged, have it replaced
by Leica service. Original power supplies are available from your Leica
branch office or Leica dealer.
Do not repair the power supply.
Electrical work must only be carried out by Leica Service.
To avoid injury to the user and for cooling reasons and fire
protection, never remove the covers of the components.
MATEO TL Microscope User Manual
this instrument is the connector between the power cable and device
port. An unobstructed access to the main circuit breaker must be
ensured at all times.
an outlet equipped with a ground contact. Do not interfere with the
grounding function by using an extension cord without a ground wire.
Any interruption of the ground wire inside or outside of the device,
or release of the ground wire connection, may cause the device to be
hazardous. Intentional ground interruption is not permitted!
Do not use this instrument near sources of high
electromagnetic radiation (for example, unshielded, intentionally
operated ultra-high frequency sources); otherwise, the proper operation
may be disrupted.
We recommend assessing the electromagnetic environment before
operation of the components and then giving corresponding
instructions.
Important Notes
By definition, the main circuit breaker of
The power plug may only be plugged into
6