Disabled:
If "Quick save image" is disabled, you need to
name a captured image before saving it.
After taking an image, the system prompts a
dialog window.
Enter the image name, and click "Confirm" to
save the image. Then the system goes back to
live view automatically.
MATEO TL Microscope User Manual
Image Format
Options: JPEG and TIFF
After system reboot, it goes back to the
default value "JPEG".
Objective Setting
Objective "Obj. HI PLAN 4x /0.10 PH0" and
objective 'Obj. HI PLAN I 10x /0.22 PH1" are
pre-installed and pre-configured before
delivery.
If an optional objective is installed,
configuring that objective in the system is
required.
Objective configuration procedure:
1. In the live view, click
to open the
system "Settings" menu.
2. Turn the knurled ring of the nosepiece to
make the target objective present at the
right of the nosepiece.
3. Click "Objective setting", a window
appears, allowing you to configure the
objective.
Preparation
The objective you are configuring in
the system is always the one at the right of
the nosepiece.
Objective setting
Click the "+" button to select the objective
1
you want to add to the current position
Then install the corresponding objective
2
to the current position
Current position is always the rightmost
hole of the nosepiece
4X
10X
Please rotate the nosepiece to the next
3
position and repeat the Step1 and 2
The system comes with the settings
for each preinstalled objective and optional
objective.
4. If the newly installed objective is one
of the two optional objectives, click the
corresponding icon of the objective to
configure it.
Contrast method:
BF, PH0
Material number:
39482938484090
4X
10X
20X
40X