Problem
The microscope does not start after pressing the On/
Off button.
The live image is too dark
MATEO TL Microscope User Manual
Possible Cause/Solution
Possible cause: No power supply.
Solution:
1. Ensure that the power outlet has power.
2. Check cable connections between:
●
microscope and power adapter
●
power adaptor and power cord
●
power cord and power outlet
Possible cause: Wrong aperture position
Solution:
●
In PH mode, adjust the aperture to the maximum.
●
In BF mode, make the aperture larger.
For details, please refer to "Setting up Contrast Mode" on page 30.
Possible cause: Too low light intensity
Solution: Adjust light intensity to make the image brighter.
For details, please refer to "Managing Camera" on page 32.
Possible cause: Mismatch of the objective and phase turret position
Solution: Change to the right phase turret position that matches the objective.
For details, please refer to "Setting up Contrast Mode" on page 30.
Troubleshooting
53