MICROPIEZO
Use a disposable syringe to withdraw and
inject the enzymatic cleaning solution
into the hard-to-reach areas (through-
holes/cannulae). Repeat this operation
three times to ensure the effective
removal of the soil from the internal
surfaces of the through-hole.;
Rinse the internal channel of the insert
with demineralised water at a pressure of
3.8 bar for at least 10 seconds so as to
eliminate any soil residue.;
Place the insert in the ultrasonic tank
immersed in the enzymatic detergent
solution at 24 °C ±2 °C and run a cycle
for at least 20 minutes;
Remove the insert(s) from the ultrasonic
tank and rinse them in demineralised
water;
50
15
x3
16
x10 sec.
H₂O
17
18