MICROPIEZO
Fill the irrigation solution tank with 300ml
of demineralised water;
NOTE: Use of distilled water is
recommended for long periods of
non-use. If the device is used after the
disinfection procedure, use common
running water.
Holding the tank in a vertical position,
insert it on the machine core unit of the
device until it comes into contact;
CAUTION: Do not overturn the tank.
The plug is not watertight. The outflow
of physiological solution or aggressive
liquids may cause damage to the
surfaces.
On the touch keyboard (see
page 17
), select:
• Power Level: 0;
• Irrigation Level: 6.
Position the scaler handpiece connector
above a container or sink to catch the
liquid released during the disinfection
cycle (see point 6 of the procedure);
Press the pedal to start the selected
program until the tank is completely
drained of all the water;
Disconnect the irrigation solution tank
from the device (see point 1) and lift the
cap;
Dry the valve area and tank with a soft,
low-lint cloth.
44
18
19
20
Chapter 5 on