Brush the internal and external surfaces
of the insert with a clean brush with soft
nylon bristles, under running water;
Rinse the insert(s) in demineralised water;
Rinse the internal channel of the insert
with demineralised water at a pressure of
3.8 bar for at least 10 seconds so as to
eliminate any soil residue.
8.5.1.4
Inserts Tightening Wrench
Place the torque wrench in a clean
container, in a horizontal position. Add
enough enzymatic detergent solution to
completely cover the torque wrench.
Leave the torque wrench to soak in
the enzymatic detergent solution for
10 minutes at ambient temperature
(22 °C ±2 °C). This procedure reduces the
amount of blood, protein, and mucus on
the instrument(s);
CLEANING AND STERILISATION