Holding the tank in a vertical position,
insert it on the machine core unit of the
device until it comes into contact;
CAUTION: Do not overturn the tank.
The plug is not watertight. The outflow
of physiological solution or aggressive
liquids may cause damage to the
surfaces.
Remove the handpiece connector from
its seat.
Position the scaler handpiece connector
above a container or sink to catch the
liquid released during the disinfection
cycle;
Press the
button on the touch
FLUSH
keyboard twice (LONG FLUSH - see
Chapter 6 on page 31
). The water and
power flow rate control will illuminate.
The button
will flash;
FLUSH
NOTE: It is possible to exit from the
cleaning mode at any time by pressing
again the button
FLUSH
will be active again by returning the last
configuration used;
CLEANING AND STERILISATION
4
5
6
7
; the keyboard
41