Mectron MICROPIEZO Mode D'emploi Et D'entretien page 142

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MICROPIEZO
Remove the handpiece, nose cone, and
light pipe from the enzymatic solution
and gently brush their surfaces using a
brush with soft nylon bristles under hot
(40 °C ±5 °C) running water, for at least
10 minutes;
Place the scaler handpiece, nose cone,
and light pipe in the ultrasonic tank
immersed in demineralised water at
40 °C ±5 °C and run a cycle for at least
10 minutes;
Rinse the inner channel of the scaler
handpiece using
a 20
previously filled with demineralised
water. Repeat five times;
Dry the surface of the scaler handpiece,
nose cone, and light pipe with a clean,
non-abrasive, low-lint cloth.
48