Mectron MICROPIEZO Mode D'emploi Et D'entretien page 141

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Rinse the inner channel of the scaler
handpiece using
a 20
previously filled with a new enzymatic
solution. Repeat twice;
Remove the handpiece, nose cone, and
light pipe from the enzymatic solution
and gently brush their surfaces using a
brush with soft nylon bristles under hot
(40 °C ±5 °C) running water, for at least
10 minutes;
Fill the ultrasonic tank with the enzymatic
detergent solution prepared according to
the manufacturer's instructions.
Use water at ambient temperature (22 °C
±2 °C);
Place the scaler handpiece, nose cone,
and light pipe in the ultrasonic tank
immersed in the enzymatic detergent
solution at 22 °C ±2 °C and run a cycle
for at least 10 minutes;
CLEANING AND STERILISATION
4
ml
syringe
5
6
7
x2
47

Publicité

Table des Matières
loading

Table des Matières