Mectron MICROPIEZO Mode D'emploi Et D'entretien page 110

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MICROPIEZO
Remove the cap of the tank and fill it with
the desired liquid;
CAUTION: The tank can hold liquids
up to a maximum of 500 ml.
The tubing and all parts of the irrigation
circuit
are
made
compatible
with
the
solutions and liquids generally used
for the treatments forming part of the
intended use of the device, specifically:
• Water (distilled, demineralised,
potable);
• Saline solution at 0.9%;
• Chlorexidine <0.3%;
• Ethyl Alcohol <15%;
• Citric Acid <5%.
CAUTION: To avoid chemical
reactions that may damage the device,
run a Long FLUSH cycle before using a
different substance (see
page 31
).
Holding the tank in a vertical position,
insert it on the machine core unit of the
device until it comes into contact;
CAUTION: Do not overturn the tank.
The plug is not watertight. The outflow
of physiological solution or aggressive
liquids may cause damage to the
surfaces.
Correctly insert the scaler handpiece
on its cord by aligning the dot on the
connector of the handpiece with the
groove on the connector of the cord.
Check that the electrical contacts of both
are perfectly dry and if necessary dry
them by blowing with compressed air;
16
from
a
material
main
clinical
Chapter 6 on
5
6
7

Publicité

Table des Matières
loading

Table des Matières