CAUTION: DO NOT USE as disinfectants::
• very alkaline products (pH > 9);
• Products containing sodium
hypochlorite;
• Products containing hydrogen peroxide;
• Products containing abrasive substances;
• Very acid products (PH < 4);
• Products containing aldehyde, amine
and/or phenols;
8.1

Preparation

1.
Perform the
FLUSH
Chapter 6 on page 31
2.
Check that all of the following
accessories have been removed/
disconnected from the machine core
unit (see
Chapter 7 on page 33
• Power supply cable;
• Pedal;
• Scaler handpiece;
• Inserts.
CAUTION: The cleaning and sterilisation
operations described in the following
chapters must be carried out at the time of
first use and at all subsequent uses.
8.2
Manual Cleaning with Enzymatic Detergent
The following procedure must be performed
on all non-sterilisable parts of the appliance,
except for the tank and its cap. The parts in
question are:
• Machine core unit;
8.2.1
Required Materials
• Clean, soft, low-lint cloths;
• Cleaning solution (pH 6-9).
36
function (see
);
):
• Acetone;
• Methylethylketone;
since they may discolour and/or damage the
plastic materials.
The manufacturer MECTRON assumes no
responsibility about damages caused by
the substances mentioned above. In case
of damage caused by those substances, the
Warranty will be void.
WARNING: Always switch off the
appliance by pressing and holding the Power
button for 2 seconds and disconnect it from
the mains before carrying out any cleaning.
CAUTION: Always disconnect the insert
from the handpiece before cleaning and
sterilising it.
CAUTION: Do not immerse the
handpiece in disinfecting solutions or other
liquids as these could damage it.
CAUTION: Do not immerse the
handpiece in the ultrasonic tank.
• Pedal and relative connection cable to
the machine;
• End part of the cord.