MICROPIEZO
The cycle can be interrupted as soon
as the liquid exits the handpiece by
pressing again the FLUSH button or,
alternatively, when the pedal is pressed
again.
Both in the case where the procedure
is interrupted and when the cycle
successfully concluded, the keyboard
will be active again by returning the last
configuration used.
Screw in the preselected insert on the
MICROPIEZO handpiece until it comes
into contact;
Insert the insert inside the torque
wrench MECTRON.
Firmly grip the body of the handpiece;
CAUTION: Do not grab the
handpiece by the nozzle and/or by
the cord. Firmly hold the body of the
handpiece and rotate only the torque
wrench. Do not rotate the body during
tightening.
Turn the torque wrench in a clockwise
direction until the notch clicks (the
external body of the torque wrench
turns with respect to the body of the
handpiece, emitting mechanical sounds).
The insert is now optimally locked;
28
4
5
6
7
click!