Cleaning Procedure - Mectron MICROPIEZO Mode D'emploi Et D'entretien

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MICROPIEZO
8.5.2.2
Cleaning Procedure
Position the accessories in a metal
basket. Connect the relevant adapter
(supplied as an optional) to the scaler
handpiece connector and then to the
water jet cleaning connections of the
thermal-disinfector.
Repeat the same operation for the
inserts by connecting them to the
appropriate adapters supplied as
optionals.
Sequence and parameters applicable to
the cycle:
• 1 min, rinse with cold water;
• 5 min, wash with alkaline
detergent
at 55°C ±2°C;
b)
• 1 min, neutralisation with
neutralising liquid
and 2/3 hot water;
• 1 min, rinse with water (1/3 cold
water, 2/3 hot water);
• 5 min, thermal disinfection at 93°C
with demineralised water.
Thermal
disinfection
tested experimentally. In compliance
with ISO 15883-1, Table B.1 [4] thermal
disinfection at a temperature of 90°C for
5 min determines a value of A0 3000.
54
, 1/3 cold water
b)
has
not
been
1
2

Publicité

Table des Matières
loading

Table des Matières