Note
Only clean the outside of the shortening sleeves with a suitable cleaning
brush.
Do not clean the inside of the shortening sleeves with a brush.
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Place ELAN 4 shortening sleeves with the larger open side facing
upward on the ECCOS holder GB727R 5, see fold-out page.
Connect the ECCOS holder GB727R 5 to the rinsing connection of the
rinsing cart.
7.8

Inspection, maintenance and checks

Allow the ELAN 4 shortening sleeves and ELAN 4 lateral burr guard to
cool to room temperature.
Check the ELAN 4 shortening sleeves and ELAN 4 lateral burr protec-
tion after every cleaning and disinfection process to ensure they are:
clean, functional, and undamaged.
Set aside the product if it is damaged.
7.9

Packaging

Always follow the instructions for use of the respective packaging and
storage devices (e.g. instructions for use TA009721 for the
Aesculap ECCOS holding system).
Place ELAN 4 shortening sleeves with the larger open side facing
upward on the ECCOS holder GB727R 5, see figure on front page.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Clean the ELAN 4 shortening sleeves and ELAN 4 lateral burr guard
under running water until all visible residues have been removed from
the surfaces.
If necessary, use a suitable cleaning brush for outer surfaces.
Chemicals
-
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
0.5 % working solution
– pH ~ 11*
-
-
In accordance with the program for the cleaning and disinfecting machine
Pack trays appropriately for the intended sterilization process (e.g. in
Aesculap sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.

7.10 Steam sterilization

Make certain that all external and internal surfaces of the product will
be exposed to the sterilizing agent.
Validated sterilization process:
– Steam sterilization using fractional vacuum process
– Steam sterilizer according to DIN EN 285 and validated according
to DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
When sterilizing several instruments at the same time in a steam ster-
ilizer: ensure that the maximum load capacity of the steam sterilizer
specified by the manufacturer is not exceeded.
9