Microflow Plus / Metering Delay; Microflow Plus / Calibration; Important; 10.3.15 Configuration / Batch - Ecolab Elados EMP II E60 Mode D'emploi

Masquer les pouces Voir aussi pour Elados EMP II E60:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10.3.14.6
MicroFlow
Use the "Metering delay" option to specify the time during which the metering monitoring
function is disabled when handling outgassing media to allow time for degassing the system.
Once the metering pause time has expired, the "Metering delay" is activated as soon as the
next metering is initiated (see chapter 10.3.14.5).
Example: Metering pause = 240 s, metering delay = 40 s
This example has a metering pause of 20 seconds during which metering is stopped. The
metering delay of 40 seconds is initiated with the next metering request during which the
metering monitoring of the
10.3.14.7
MicroFlow
Use the "Calibration" option to newly calibrate the transducer during initial installation or when
changing the product.

IMPORTANT

On activating the "calibration" function, the values for min. conductivity (electrical connection
between the sensor pegs in the measuring sensor disrupted) und max. conductivity (electrical
connection) are determined.

10.3.15 Configuration / Batch

NOTE
When activating batch mode with a signal to the metering lock/pulse input (plug II) (or at plug
VII if a dongle box is used) pin
defined quantity is metered with 100 % stroke frequency (assortment of inputs see chapter
10.3.15.3, electrical connection see chapter
The batch metering can be interrupted by deactivating the release (metering lock) or
switching off the pump.
417101438_EMP_II_E60
/ Metering delay
PLUS
MicroFlow
/ Calibration
PLUS
The pump must be running and the metering medium must flow through the transducers
without any bubbles before starting the calibration process.
The pump must be calibrated before using batch metering
(see chapter:
10.4
"Calibration").
2+4
PLUS
is interrupted.
These values are then used to calculate and
display a difference.
To monitor the system well, the difference should
be greater than 3..4.
Fig. 10.46 Configuration MicroFlow
(input pulse) or pin
3+4
7.2.3.4
and
8.2.3.1
- 116 -
PLUS
Fig. 10.45 Configuration MicroFlow
Any gas bubbles can be removed
from the system without triggering
an alarm during this time.
Setting range: 2 - 3600s
Default setting = 40 sec.
PLUS
/Calibration
(input metering lock), a previously
"Installing the batch pulse").
/ Metering delay
Rev. 12-01.19