11.0  ERROR
CODES 
 
During the initial power‐on testing or during operation, the scale may show an error 
message.  The meaning of the error messages is described below. 
If an error message is shown, repeat the step that caused the message, turning the 
balance on, carry out the calibration or other functions. If the error message is still 
shown contact your dealer for further support.   
ERROR CODE 
Err 1
Err 2
Err 4
Err 6
diSPEr
DESCRIPTION 
Time input error.
Date input error
Initial  Zero  is  greater 
than  allowed  (typically 
4%  of  the  maximum 
capacity) when power is 
turned  on  or  when  the 
[Zero] key is pressed, 
A/D count is not correct 
when  turning  the  scale 
on. 
Shows  when  a  number 
is  too  large  tot  fit  the 
display 
POSSIBLE CAUSES
Tried  to  set  an  illegal  time,  i.e. 
26hours  
Tried to set an illegal date, i.e. 36th 
day  
Weight  is  on  the  pan  when  turning 
the scale on. 
Excessive  weight  on  the  pan  when 
zeroing the scale. 
Improper calibration of the scale. 
Damaged load cell. 
Damaged Electronics. 
Platform is not installed. 
Damaged Load cell. 
Damaged Electronics. 
 
The accumulated totals for weight or 
count  exceed  999999.    The  data  on 
RS232 is still shown accurately 