Inserting a Specimen Mover
Plate
Aligning the TegraForce-5
TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual
Use the Raise/Lower ON button to ensure that TegraForce-5 is
fully raised.
Push the black button on the TegraForce-5.
Insert the specimen mover plate and rotate it until the three pins
are aligned and then push the specimen mover plate upwards
until it locks in position.
Release the black button on the TegraForce-5 and ensure that
the specimen mover plate is securely fixed.
TegraForce-5 is equipped with an automatic alignment mechanism
taking care of the fine alignment of the head relative to the
preparation disc. It ensures that when the TegraForce-5 head is
lowered, it returns to its previous position. To change the alignment
of the head:
With a specimen mover plate or specimen holder fitted, press
the Raise/Lower button to lower the head.
Push the entire head sideways (left or right) until it is aligned
correctly. The specimens should be positioned as close to the
edge of the preparation disc as possible, in fact they may extent
over the edge by 2-3 mm.
To adjust the friction of the sideways movement see: Adjusting
the friction in the section: Setting up TegraForce-5 on TegraPol
The automatic alignment is only a fine alignment of the TegraForce-5. It has
to be positioned in approximately the correct position, max. ± 20° from the
IMPORTANT!
correct position.
Black button
Pins