Daily Service
Weekly Service
TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual
10. Maintenance
Clean all accessible surfaces with a moist cloth.
Check and refill TegraDoser-1.
Remove the grinding/polishing disc and remove all dirt from the
tray and the drain tube.
Clean the pressure feet and pistons applying the force on the
specimens and specimen holder.
Press the release outlet valve on TegraForce to drain the
water/oil filter (please refer to the section on
/ Oil
Filter).
e
From the Main Menu, select Configuration and press
e
ENTER
to activate the CONFIGURATION Menu.
D
D
Select Cleaning and press START
s
pressure feet or the centre piston down.
D
Clean the pressure feet and the pistons or the centre
piston using a dry cloth.
o
Press STOP
o
piston up again.
e
to move the pressure feet or the centre
Emptying the Water
to move the
55