Checking the Packing
Contents
Getting Acquainted with
TegraDoser-1

Filling TegraDoser-1

TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual
7. Operating TegraPol
with TegraForce and TegraDoser-1
1
TegraDoser-1
1
Funnel
1
Flat blade screwdriver
1
Sheet with labels
The TegraDoser bottle has a snap tight lid, which provides easy
access when filling the bottle.
Ensure that no preparation process is active.
Open the snap lid on the top of the bottle and insert the funnel
supplied with the TegraDoser.
Fill/refill the bottle with lubricant / suspension and snap the top
closed.
TegraDoser-1 can only be used with lubricants and diamond suspensions.
It is perfectly suited for use with the new all-in-one DiaPro Suspensions
allowing unattended preparation.
It cannot be used with oxide polishing suspensions. Oxide polishing
suspensions are to be used on TegraDoser-5.
Lid
Rotary switch for ID setting
Dosing nozzle
Important!