Setting up TegraForce-3 on
TegraPol
22
TegraPol-21/ -25/ -31/ -35
TegraForce-3/ -5
Instruction Manual
Remove the cap covering the support hole placed on the top of
the TegraPol cabinet.
Support Hole
Fastening
Screws
Remove the two caps covering the fastening screws on the left-
hand side of the TegraPol cabinet by lifting them off with a knife.
Lead the supporting column of TegraForce-3 down into the
support hole in TegraPol.
Turn the column until the cross pin at the end of the column
engages in the slot in the casting of the TegraPol while the
specimen mover is placed over the preparation disc.
Supporting
Column
Fastening
Screws
Tighten the two fastening screws on the left-hand side of the
TegraPol cabinet and mount the two caps again.
If it is not possible to swing the TegraForce-3 over the left hand
side of the machine when a sample holder is mounted, loosen
the two fastening screws, lift the specimen mover slightly up to
allow the TegraForce to swing freely and tighten the fastening
screws again.