Placing the Specimen

Pouring Resin over the Specimen

Placing Two Specimens

10
LaboPress-3
Instruction Manual
Press and hold RAM UP
limit.
Apply "Mould Release Agent" to the surface of the lower ram.
Place the specimen on the ram. The specimen must be clean,
dry and free from grease. The distance between the specimen
and the cylinder wall must be minimum 3 mm to avoid cracks
in the resin.
Press and hold RAM DOWN
limit.
Fill a suitable amount of resin into the cylinder by means of
the enclosed funnel.
Always make sure there is sufficient resin to cover the sample
after compression. Please note that the volume of the resin
reduces when the granulate becomes compressed. If insufficient
resin is used, the rams may come in contact with the sample, and
the cylinder may be damaged.
The "Mould Release Agent" must always be applied to the mounting rams
as a thin layer to prevent the resins sticking to the surface. By means of
Struers AntiStick a thin layer of stearate powder can easily be dabbed on
Follow the instructions in Placing the Specimen.
Apply "Mould Release Agent" to all surfaces of the
intermediate ram.
Place the intermediate ram on top of the resin.
Place the second specimen on the intermediate ram and fill
the cylinder with an adequate quantity of resin.
Again, make sure you use adequate resin!
O
to raise the lower ram to its upper
N
to lower the ram to its lower
Important
the rams.