BED EXIT SYSTEM (OPTIONAL)

WARNING
The Bed Exit system (optional) is intended only to help in the detection of a patient exiting the bed . It is not intended to
replace patient monitoring protocol . The Bed Exit system signals when a patient is about to exit the bed . The addition or
removal of equipment with a Bed Exit system armed must be done using the "Adding or Removing Equipment with the
System Armed" procedure (on
the patient's movement in the bed be erroneous .
The Bed Exit system (optional) is not designed to be used with patients weighing less than 50 lb (23 kg) .
K:
Press to activate or deactivate the Bed Exit .
M:
Status LED .
Note
For the Bed Exit alarm signal to be heard in the nurse station, the bed must be equipped with the optional nurse call
function . For beds without the optional nurse call, the alarm signal will be heard only in the patient's room .
BED EXIT
Zeroing the System
The Bed Exit system absolutely must be zeroed before use:
• Zero the scale if not already done (see "Zeroing the Scale System" on
Arming the System
Following the zeroing of the system, the Bed Exit system may now be activated:
• Install the patient in the bed . Ensure the patient lies in the bed in its usual position before activating the system .
• Activate the Bed Exit by pressing Arm/Disarm (K) . The status LED (M) will light up .
• To deactivate a Bed Exit alarm or to turn the system off, press Arm/Disarm (K) . The status LED (M) will turn off .
Adding or Removing Equipment with the System Armed
Changing equipment on a bed with the Bed Exit system armed requires the user to follow either procedure described
below:
• If the bed is equipped with the scale system, follow the "Adding or Removing Equipment when a Patient is on the
Bed" on
page
1-21 .
-or-
• If the bed is not equipped with the scale system, remove the patient from the bed; change the equipment and zero
the system (see ¨Zeroing the System¨ on
to the bed .
Return To Table of Contents
1-22
Operation Guide
page
1-24), otherwise the sensitivity of the system may be affected and the readings of
page
1-23) . The Bed Exit system may be activated once the patient returns
28EX-009-001 REV B