Fixed Three Stage I.v. Pole, Foldable - Fa64170 - Stryker GoBed II Lit MedSurg FL28EX Guide D'utilisation

Lit d'hôpital ajustable
Masquer les pouces Voir aussi pour GoBed II Lit MedSurg FL28EX:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Fixed Three Stage I.V. Pole, Foldable - FA64170

The fixed three stage I . V . pole FA64170 is a permanently attached, telescoping I . V . pole . The extension of the telescoping
section provides several additional height positions . The I . V . pole can fold into a storage position .
WARNING
The scale and bed exit systems must be adjusted if this accessory is added when either system is in operation . Refer to
the "Add / Remove Equipment" procedure
Tools Required:
• Ratchet with a 7/16" and 1/2" Socket
• 7/16" and 1/2" Combination Wrench
• #2 Phillips Screwdriver
I.V. POLE INSTALLATION
Note: The I . V . pole can only be installed on the right side of the
head end of the bed .
1 .
Remove the head board .
2 .
Using a 1/2" combination wrench, remove the bolt (A) and
spring washer (B) from the base of the I . V . pole .
3 .
Install the I . V . pole into the socket and attach it using the bolt
and spring washer removed in step 2 .
Note: The bolt is coated with "Scotch Grip" . The coating becomes
less effective once the bolt is tightened and removed thereafter .
Replace the bolt if the I . V . pole is removed and reinstalled in an-
other socket or bed .
4 .
Using a 1/2" combination wrench, remove the two nuts (C)
and bolts (D) .
5 .
Install the storage plate (G) and secure with the bolts and nuts
removed in step 4 .
6 .
Reinstall the head board .
I.V. POLE OPERATION
1 .
Lift and pivot the pole from the storage position and push down until it is locked into the receptacle .
2 .
To increase the height of the pole, pull up on the first telescoping portion (E) until it locks into place .
3 .
To increase the height further, pull up on the second telescoping portion (F) . Release the section at any desired
height and it will lock into place .
4 .
To lower the second telescoping section, push up on the red part of the handle (J) while holding the section . Lower
the section to the desired height and release the handle to lock the section into place .
5 .
Rotate the I . V . pole hangers (H) to the desired position and hang the I . V . bag(s) .
WARNING
The weight of the I . V . bags must not exceed 40 lb (18 kg) .
6 .
To store the I . V . pole, push up on the red portion of the handle (J) while holding onto the second section and lower
it . Turn the latch (K) clockwise to lower the first section . Lift the pole to remove it from the receptacle . Fold the pole
down and store it on the bracket .
Return To Table of Contents
1-40
(page
1-21) .
28EX-009-001 REV B
H
J
D
G
B
C
E
F
K
A
www.stryker.com

Publicité

Chapitres

Table des Matières
loading