Stryker GoBed II Lit MedSurg FL28EX Guide D'utilisation page 23

Lit d'hôpital ajustable
Masquer les pouces Voir aussi pour GoBed II Lit MedSurg FL28EX:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SCALE SYSTEM (CONTINUED)
Adding or Removing Equipment when a Patient is on the Bed
• Press
to activate the scale display . Access the Change Equipment mode using the Menu Down key, the display
will read:
• Press Enter . The display will read:
Do Not Touch Bed (ensure no one touches the bed while this message is displayed) .
• The following message will then appear:
Signalling that equipment can now be added or removed from the bed . When done with the equipment, press Enter .
• Display will then read Release to Fin . (release Enter), followed by:
Do Not Touch Bed (ensure that the patient remains still and that no one touches the bed while this message is
displayed) .
The system will return to the Scale mode displaying the patient's weight (the weight should be the same as the one
displayed before the procedure) .
Manually Changing the Numerical Value of Displayed Weight
In certain circumstances, it may be necessary to manually change the patient's weight displayed . To do so:
• Press
to activate the scale display . Access the Change Patient Weight mode using the Menu Down key . The
display will read:
• Press Enter . The display will read:
Do Not Touch Bed (ensure no one touches the bed while this message is displayed) .
• The display will then read  XXX .X lb / press Enter . Adjust the weight using the Menu Up/Down controls .
When the desired weight is entered, press Enter . The system will return to the Scale mode and display the modified
weight .
Note
The Change Patient Weight mode should not be used systematically for new patients . The scale should be zeroed each
time a new patient occupies the bed .
Changing the Weight Unit of Measure
• Press
to activate the scale display . Access the Select Weight Unit (lb/kg) mode using the Menu Up key . The
display will read:
• Press Enter . The display will read:
• Select the weight unit using the Menu Up or Down control and press ENTER . The system will return to the Scale
mode with the weight displayed in the unit chosen .
www.stryker.com
Operation Guide
Change Equip
Press Enter
Release to Start (release Enter), followed by:
Add/Remove Equip
Press Enter
Change Patient Weight
Press Enter
Release to Change Weight (release Enter), followed by:
Units
Press Enter
 lb and OR  kg and
Press Enter
28EX-009-001 REV B
Press Enter
Return To Table of Contents
1-21

Publicité

Chapitres

Table des Matières
loading

Table des Matières