Fixed Two Stage I.V. Pole, Foldable - FA64171
WARNING
Ensure all the connectors stay plugged into the electronic board if it is present on the head casing cover .
BUMPERS INSTALLATION
1 .
Unplug the power cord from the bed .
2 .
Using a 1/2" combination wrench with a 1/2'' socket, remove the bolts, washers and nuts holding the rollers in place .
3 .
Using a Phillips screwdriver, remove the screws (L) holding the head casing cover .
4 .
Using a 7/16" combination wrench with a 7/16" socket, remove the nuts (J) and the bolts (H) holding the bumper
brackets to the head casing cover .
5 .
Mount the new brackets (F) on the head casing cover using the new hexagonal bolts (H) and nuts (J), as shown
in the drawing below .
6 .
Install the head casing cover back on the bed frame using the screws (L) removed in step 3 .
7 .
Insert the rollers (G) on each bracket and secure using the new hexagonal bolts (I) and nuts (K) .
Note: When tightening the nuts, make sure that there is enough room for the rollers to spin freely.
L
K
www.stryker.com
F
H
G
28EX-009-001 REV B
J
I
Return To Table of Contents