The overhead trapeze patient helper provides a support that allows the patient to change position in bed . The system
features a fixed lower section, a mobile upper section and a trapeze . When in use, the upper section is centered over
the patient . When not in use, the upper section can rotated to either side of the bed for storing .
WARNING
The scale and bed exit systems must be adjusted if this accessory is added when either system is in operation . Refer to
the "Add / Remove Equipment" procedure
INSTALLATION
Tools Required:
• None
Procedure:
1 .
Insert the lower section posts (A) into the slots on the head
end of the bed . Ensure the posts seat completely into the
slots
2 .
Insert the hitch pins (B) into the holes (C) located on the ends
of each post .
3 .
Place the nylon washer (D) onto the pivot pin (E) of the lower
section .
4 .
Pull the lock pin (G) and install the upper section onto the
pivot pin (E) of the lower section .
5 .
Verify the lock pin (F) on the upper section is seated in an
adjustment hole .
OPERATION
WARNING
• The overhead trapeze is designed to help the patient change
position in bed . The trapeze must not be used as a support to
enter or exit the bed . Injuries may result due to improper use
of the overhead trapeze . Medical advice may be necessary
to determine if the use of an overhead trapeze is adequate
for the patient's condition or treatment given .
• The maximum load capacity of the overhead trapeze is 150
lb (68 kg) .
1 .
To position the trapeze over the patient, pull the lock pin (F)
and pivot the upper section towards the center of the bed .
Release the lock pin and ensure it is properly engaged .
2 .
To adjust the trapeze (H) height, shorten or lengthen the
strap (J) . If you must detach the strap to free the trapeze,
ensure the strap is properly reinserted in the fastener .
Return To Table of Contents
1-30
Patient Helper - FA64148L
(page 1-21) .
28EX-009-001 REV B
K
G
F
D
E
A
C
C
B
B
www.stryker.com
J
H