Update Leads - Boston Scientific Vercise Neural Navigator Mode D'emploi

Table des Matières

Publicité

Update Leads

This feature allows you to enable supported leads within the Leads Configuration Panel. The
Update Leads feature can be accessed from the Tools Tab. The DB2201-Standard Lead is
enabled by default.
Note: You can only update leads when disconnected from the Stimulator.
To enable the DB2202-Directional lead (See Figure 10):
1. Disconnect from the patient's Stimulator by clicking on the Disconnect Button in the
Connect Tab.
2. Select the Tools Tab.
3. Select the Update Leads Tab.
4. Select the row for the desired lead.
5. Select the column labeled Key and enter the key 'KEVHSK'.
6. Select Verify Key(s).
Figure 10. Update Leads
Vercise™ Neural Navigator Programming Manual
Tools and Additional Information
91046489-05
17 of 557