Vercise™ Neural Navigator Programming Manual

Charge Density

Figure 12. Charge Density Limits for Boston Scientific DBS Leads
Figure 12 displays the recommended maximum charge density for different combinations of
Amplitude (mA) and Pulse Width (μs). The solid black line (Limit: 6mm
DB2201-Standard Lead and the proximal and distal contacts of DB2202-Directional Lead. The
dashed black line (Limit: 1.5mm
Lead. These estimates of charge density are only for the Boston Scientific DBS Leads
WARNING: Patients may have the ability to change the Amplitude with the Remote
Control.The physician should set and verify the maximum and minimum
Amplitude levels allowed by the Remote Control to ensure that current levels
remain safe.
Vercise™ Neural Navigator Programming Manual
91046489-05
20 of 557
Stimulation Limits
Single Electrode
Pulsewidth [
) refers to the small directional contacts of DB2202-Directional
2
Limit: 6 mm
Limit: 1.5 mm
μ
s]
) refers to all contacts on
2
2
2