Introduction
Vercise™ Neural Navigator is a software program that is used to set and adjust stimulation
parameters for the Vercise PC DBS System.
An initial programming session may consist of the following steps:
1. Starting Vercise Neural Navigator.
2. Connecting to the Stimulator.
3. Configuring the Stimulator and Leads.
4. Testing different Stimulation Settings.
This manual will provide instruction on how to accomplish these steps as well as perform additional
functions, such as exporting reports and backing up data.
If you have any issues, please contact Boston Scientific Customer Service.
Note: Screens depicted in this manual may differ slightly from the screens on your Vercise
Neural Navigator.
Intended Use
Vercise Neural Navigator is a software program that is used to set and adjust stimulation
parameters for the Vercise PC DBS System.
Connecting the Programming Wand to the Clinician Programmer
The Clinician Programmer (CP) communicates with the Stimulator via a Programming Wand
(See Figure 1). The Programming Wand uses a radio frequency (RF) link to communicate with
the Stimulator.
Figure 1. Clinician Programmer (CP) and Programming Wand
Vercise™ Neural Navigator Programming Manual
91046489-05
Introduction
1 of 557