Vercise™ DBS Leads Directions for Use
(e). Create a tunnel, for the proximal end of the DBS Lead, to a place closer to the desired
DBS Extension connector location.
CAUTION: Placement of the DBS Extension connector in the neck region can increase the
risk of device failure due to repetitive movement of the neck.
(f). Create a pocket under the skin for the excess DBS Lead and Lead Boot.
(g). Coil excess DBS Lead material under the scalp, in the pocket, until it is ready to be
connected to the DBS Extension.
Note: The DBS Lead may be connected to the DBS Extension and Stimulator during a separate
surgery at a later time. See "Stimulator Implantation" in the appropriate IPG DFU as listed
on your DBS Reference Guide.
7. Repeat the "Implanting the DBS Lead" procedure for the second DBS Lead. Use the Tunneling
Tool to tunnel the second DBS Lead to the same side as the first Lead.
8. Close the incisions.

Explanting the DBS Leads

For instructions on explanting the DBS leads, see the appropriate DFU for your Boston Scientific
DBS System as listed on your DBS Reference Guide.
Vercise™ DBS Leads Directions for Use