4. Operation

Control handset

The handset gives the user control over backrest angle, leg elevation, bed
height and tilt angle (model 161 only). The Auto-Contour function allows easy
simultaneous adjustment of leg elevation and backrest angle.
When Auto-Contour is used to flatten the mattress platform, the leg
sections remain raised until the backrest has partly lowered. This
helps to prevent the patient slipping down the bed.
To operate a function, press and hold the appropriate button until the desired
position is obtained.
Auto Contour – adjusts the backrest and leg sections to
achieve a comfortable sitting position.
Bed Height – adjusts the height of the mattress platform
above the floor.
Backrest – adjusts the angle of the backrest section.
Leg Elevation – adjusts the height of the calf section
above the mattress platform.
Tilt (model 161 only) – allows the use of head down or
foot down tilt.
The bed cannot be used in applications that need the patient to be put
in the Trendelenburg position. The Maximum tilt angle of the Minuet
bed frame is 10 .
The lockable handset has the facility to selectively disable bed functions to
prevent use by the patient.
Between each pair of pushbuttons is a small rotary lockout switch.
To disable a function: fit the plastic key into the two holes in the lockout switch
(A) and turn the switch clockwise to the Locked position (B).
To enable a function: fit the plastic key into the two holes in the lockout switch
and turn the switch anti-clockwise to the Unlocked position (C).
To enable/disable a function the plastic key must be used.
Only use the plastic key supplied. Using other tools may damage the
switch.
WARNING
A
Caution
B
C