7
K
Plug the curly cable (I) into the head end height actuator (J). Make sure
the plug is fully inserted in the socket. Push in the cable retaining clip (K)
until it "clicks" into place.
Repeat this procedure for the foot end height actuator.
L
8
Squeeze together the split ends of the retaining pins (L) and (M), and
partially push them out of the head end cable channel (N) as shown,
above left. Do not allow the cables to spring out of the channel.
Carefully straighten out the cable(s) from the foot end section, and
position them in the head end cable channel alongside the other cables.
Refit the two retaining pins to hold all the cables in place within the head
end cable channel.
I
J
N
M
17