Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601-1-2
Immunity test
Electrostatic
±2kV, ±4kV,
discharge (ESD)
±8kV, ±15kV air
EN 61000-4-2
±8kV contact
3V in 0,15 MHz
to 80 MHz
Conducted
disturbances
6V in ISM and
inducted by RF
amateur radio
fields
bands between
0,15 MHz and
EN 61000-4-6
80 MHz
80% AM at 1
kHz
Home
Healthcare
Radiated RF
environment
electromagnetic
10 V/m
field
80 MHz to 2,7
GHz
EN 61000-4-3
80% AM at 1
kHz
Proximity fields
385 MHz - 27
from RF wireless
V/m
communications
450 MHz - 28
equipment
V/m
EN 61000-4-3
710, 745, 780
MHz - 9V/m
810, 870, 930
MHz - 28 V/m
1720, 1845,
1970, 2450 MHz
– 28 V/m
5240,5500,
5785 MHz -
9V/m
±1kV SIP/SOP
ports
Electrical fast
±2kV AC port
transient/burst
100kHz
EN 61000-4-4
repetition
frequency
Power frequency
30A/m
Magnetic field
50 Hz or 60 Hz
EN 61000-4-8
Compliance
test level
level
±2kV, ±4kV,
±8kV, ±15kV air
±8kV contact
3V in 0,15 MHz to
80 MHz
6V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80% AM at 1 kHz
Home Healthcare
environment
10 V/m
80 MHz to 2,7
GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780
MHz - 9V/m
810, 870, 930
MHz - 28 V/m
1720, 1845, 1970,
2450 MHz - 28
V/m
5240,5500, 5785
MHz - 9V/m
±1kV SIP/SOP
ports
±2kV AC port
100kHz repetition
frequency
30A/m
50 Hz
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material the relative humidity
level should be at least 30%.
Portable and mobile RF communications
equipment should be used no closer to
any part of the product, including cables,
than 1.0m, if the transmitter's output
power rating exceeds 1W
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
b
frequency range
.
Interference may occur in the vicinity of
equipment marked with this symbol:
Mains power supply should be that of a
typical commercial or hospital
environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.