Output Power Diagrams - Medtronic Prostiva RF 8930 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 127
8930EN_6.fm 12/17/10 2:44 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
Chapter 6
96
Device Specifications
English 8930 System User Guide 2010-12
MA07445A008
Rev X

Output Power Diagrams

Power (Watts)
Power (Watts)
16
15
14
12
10
8
6
4
2
35 100
200
Figure 6-1. Output Power vs. Load
200
180
160
140
120
100
80
60
40
20
0
0
2
4
6
8
10
Figure 6-2. Output Voltage vs. Power Setting
Note: RF Generator is rated at 50 W, however, software algorithm
limits maximum output to 15 W or 95 Vrms (each channel).
300
500
600
400
Impedance (Ohms)
Impedance (Ohms)
12
15
20
25
30
35
40
Power Setting (W)
Power Setting (W)
Medtronic Confidential
CRMRef_R06
600 Ohms
600 Ohms
500 Ohms
500 Ohms
400 Ohms
400 Ohms
300 Ohms
300 Ohms
200 Ohms
200 Ohms
100 Ohms
100 Ohms
50 Ohms
50 Ohms
35 Ohms
35 Ohms
45
50

Publicité

Table des Matières
loading