Radio Frequency Generator Description
Table 2-8. Caution Messages (continued)
No.
Message
54
Needle impedance (Ω) low
55
Left needle impedance (Ω)
high
56
Right needle impedance
(Ω) high
57
Needle impedance (Ω)
high
User Interface
Explanation
Occurs only at start of a
lesion when both needles
have impedance
< 80 ohms
Nominal starting range is
80 to 180 ohms
Physician may continue
with lesion
Message stays on for ten
seconds
Alarm is 2 short beeps
Occurs only at start of a
lesion when left needle has
impedance >180 ohms
Nominal starting range is
80 to 180 ohms
Physician may continue
with lesion
Message stays on for ten
seconds
Alarm is 2 short beeps
Occurs only at start of a
lesion when right needle
has impedance >180 ohms
Nominal starting range is
80 to 180 ohms
Physician may continue
with lesion
Message stays on for ten
seconds
Alarm is 2 short beeps
Occurs only at start of a
lesion when both needles
have impedance
>180 ohms
Nominal starting range is
80 to 180 ohms
Physician may continue
with lesion
Message stays on for ten
seconds
Alarm is 2 short beeps
2010-12 8930 System User Guide English