Chapter 7
108
Periodic Inspection and Maintenance
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Operating test
The following should be functionally tested:
Set-up generator and turn on power to check the self-test
diagnosis and the audible tone output.
Configure the system and confirm that the radio frequency
power shuts down when the return electrode is disconnected.
Electrical safety test
Complete the following tests in accordance with the applicable
section of the international standard IEC 60601-1:
Test the protective earth resistance limit: 0.1 Ohms.
Test the earth leakage current limit: .5 mA.
Test the enclosure leakage current limit: .1 mA.
Cleaning
When necessary, clean the radio frequency generator according
to the following guidelines:
Always unplug the generator before cleaning.
Keep liquid, including the cleaning fluid, out of the generator's
case and any openings.
Do not use spray cleaner directly onto the generator.
Do not use harsh or caustic chemical products to clean the
generator.
Clean the exterior case of the generator with a soft cloth lightly
dampened with water.
To clean the LCD screen, dilute a household water based
glass cleaner to a 50/50 ratio with water and spray a small
amount on a clean, soft cloth and gently wipe the screen.