Bosch DCI 700 Notice Originale page 19

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 55
1
2
7. Attach the front left cover.
8. Check the fill level in the test oil tank after an injec‐
tor test. If necessary, top up the test oil.
---
5.2.3
Changing the test oil filter
---
5.2.3.1 Removing the test oil filter
1. End all computer applications.
2. Shut down the Windows operating system.
3. Switch off the DCI 700 at the main switch (2).
4. Remove the front left cover (1).
1
5. Release the two wing screws
move the wing screws.
Robert Bosch GmbH
3
4
2
(1)
and (4). Do not re‐
(2)
6. Use the two handles
tank out of the test bench.
1
2
7. Place a collecting container
ter.
8. Loosen the test oil filter
with the aid of a strap wrench.
9. Unscrew the test oil filter
test oil in the collecting container
10. Dispose of the test oil and test oil filter (see
"Disposal").
---
DCI 700 | 19 | en
(3)
and
to pull the test oil
3
4
(2)
under the test oil fil‐
(1)
by turning it clockwise
(1)
. Collect the draining
(2)
.
1
2
 25
1 689 989 473 | 2023-09-24