Free air delivery (FAD) correction due to elevation
Capacity given in the form of FAD („Free Air Delivery") applies to the following conditions:
Altitude
Atmospheric pressure
To calculate FAD compressor capacity in dependence on altitude, it is necessary to apply
correction factor according to the following table:
Altitude [m.n.m.]
FAD correction factor
Electromagnetic compatibility declaration
Medical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to
be installed and put into service according to the EMC information provided below.
Guidance and manufacturer's declaration - electromagnetic emissions
Pursuant to IEC 60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for
basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
The equipment is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
NP-DK50 PLUS, 2V-18_02-2023-MD
0 m.n.m.
101325 Pa
0 -1500
1
Compliance
Group 1
Class B
Class A
The equipment is not
likely to cause any flicker,
as the current flow is
approx. constant after the
start up.
- 16 -
Temperature
Relative humidity
1501 - 2500
2501 - 3500
0.80
Electromagnetic environment - guidance
The equipment uses RF energy only for its
internal functions. Therefore, the RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The equipment is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low voltage power supply
network that supplies buildings used for
domestic purposes.
DK50 PLUS, DK50 2V
20°C
0%
3501 - 4500
0.71
0.60