COMPACT PIEZO LED
9.1.3
Guide and the Manufacturer's Declaration - Electromagnetic
Immunity
9.1.3.1
Power Connection BC Input
COMPACT PIeZO leD is designed to operate in the electromagnetic environment specified
below.
The purchaser or user of COMPACT PIEZO LED should ensure that it is used in such an
environment.
Essential EMC
Phenomenon
standard or test
Electrical fast
transient/burst
IEC 61000-4-4
a) l) o)
Pulses
Differential
IEC 61000-4-5
mode
a) b) j) o)
Pulses
common
IEC 61000-4-5
mode
a) b) j) k) o)
Conductive
disturbances
IEC 61000-4-6
induced by RF
fields
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage inter-
IEC 61000-4-11
ruptions
f) i) o) r)
38
Immunity test values
method
±2 kv on contact
100 KHz repetition
frequency
± 0.5 kv, ± 1 kv
± 0.5 kv, ± 1 kv, ± 2kv
3 V
m)
0.15 MHz - 80 MHz
6 V
in the ISM bands
m)
between 0.15 MHz and
80 MHz
80 % AM at 1 KHz
0% UT; 0,5 cycle g)
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle e
70 % UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
Environment Guidance
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF com-
munication devices should
not be used near any part of
the product, including cables,
except when they respect the
recommended and calculated
n)
distances from the equation
e)
applicable at the frequency of
the transmitter.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
Electromagnetic