Rinse the internal channel of the insert
with demineralised water at a pressure of
3.8 bar for at least 10 seconds so as to
eliminate any soil residue.;
Place the insert in the ultrasonic tank im-
mersed in the enzymatic detergent solu-
tion at 24 °C ±2 °C and run a cycle for at
least 20 minutes;
Remove the insert(s) from the ultrason-
ic tank and rinse them in demineralised
water;
Brush the internal and external surfaces
of the insert with a clean brush with soft
nylon bristles, under running water;
CLEANING AND STERILISATION
16
x10 sec.
H₂O
17
18
19