COMPACT PIEZO LED
While soaking in the enzymatic solution,
gently brush all the surfaces until all visi-
ble soil has been removed.
Use a clean brush with soft nylon bristles
for the external surfaces, and a clean pipe
cleaner with soft nylon bristles for the in-
ternal cavities and grooves.
NOTE: Thoroughly brush all the
following parts of the insert for about 20
seconds:
• Through-holes and internal
channels;
• Hard-to-clean areas such as
sharp edges and in particular, the
interstices between the cutting
cusps;
• Internal cavities, grooves and
fissures.
Remove the insert from the enzymatic
detergent solution.
Thoroughly rinse and brush all the surfac-
es of the insert (see previous point) under
running water for at least 10 minutes;
Use a disposable syringe to withdraw and
inject the enzymatic cleaning solution
into the hard-to-reach areas (through-
holes/cannulae). Repeat this operation
three times to ensure the effective re-
moval of the soil from the internal surfac-
es of the through-hole.;
24
13
14
15
x3