Position the accessories in a metal basket.
Connect the relevant adapter (supplied as
an optional) to the scaler handpiece con-
nector and then to the water jet cleaning
connections of the thermal-disinfector.
Repeat the same operation for the inserts
by connecting them to the appropriate
adapters supplied as optionals.
Sequence and parameters applicable to
the cycle:
• 1 min, Rinse with cold water;
• 5 min, Wash with alkaline detergent
at 55°C ±2°C;
• 1 min, Neutralisation with suitable
solution (1/3 cold water, 2/3 hot
water);
• 1 min, Rinse with water (1/3 cold
water, 2/3 hot water);
• 5 min, Thermal disinfection at 93°C
with demineralised water.
Thermal disinfection has not been tested
experimentally. In compliance with ISO
15883-1, Table B.1 [4] thermal disinfec-
tion at a temperature of 90°C for 5 min
determines a value of A0 3000.