COMPACT PIEZO LED
9.1
Electromagnetic Compatibility IEC/EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment
Even if compliant with standard IEC/EN
60601-1-2, COMPACT PIEZO LED may
interfere with other devices in the vicinity.
WARNING: Portable and mobile radio
communication equipment may influence the
correct operation of the appliance.
9.1.1
Guide and Manufacturer's Declaration - Electromagnetic
Emissions
COMPACT PIeZO leD is designed to operate in the electromagnetic environment specified
below. The purchaser or user of COMPACT PIEZO LED should ensure that it is used in such an
environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
emissions of fluctuations
voltage/flicker
IEC 61000-3-3
36
WARNING: Contraindications.
Interference from other equipment
An electrosurgical scalpel or other
electrosurgical units positioned near the
COMPACT PIEZO LED appliance may interfere
with the correct operation of the appliance
itself.
WARNING: The appliance requires
particular EMC precautions and must be
installed and put into service according to the
EMC information provided in this chapter.
WARNING: The use of cables and
accessories not supplied by MECTRON may
adversely affect the eMC performances.
Compliance
Electromagnetic Environment Guidance
COMPACT PIEZO LED uses RF energy only
for its internal operation. Therefore, its RF
Group 1
emissions are very low and probably do not
cause any interference with nearby electron-
ic devices.
Class B
COMPACT PIEZO LED is suitable for use in
all buildings, including domestic buildings,
Class A
and those directly connected to the public
low-voltage power supply network that sup-
plies buildings used for domestic purposes.
Compliant